Fault repair of Omron blood pressure monitor

Create Date: 2024-8-30 12:03:23|Source: Omron/Omron

1. Mercury leakage

1. The cause of mercury leakage failure

Under normal circumstances, the sphygmomanometer will often leak mercury, which may eventually cause a certain error in the measurement value of the sphygmomanometer device. Then the fault is generally manifested in the switch of the sphygmomanometer device is opened, the mercury column in the tube does not reach the level of the zero scale line, and there are bubbling and broken columns in the process of pressurizing the sphygmomanometer device. In the process of verification of the sphygmomanometer device, if the mercury in the sphygmomanometer device reaches 38kpa, then the sphygmomanometer needs to be placed in a stationary state, and it needs to be observed for two minutes or more, and finally the cause of the mercury leakage can be understood, which is mainly caused by the failure of the components of the sphygmomanometer device, and there are also some human factors.

2. Mercury leakage fault maintenance

When there is mercury leakage at the lowest end of the glass tube of the sphygmomanometer device, it can be said that the sealing rubber gasket at the lower end has been aged, and the rubber pad should be replaced, if the elasticity of the upper pressure plate spring of the glass tube of the sphygmomanometer device itself is relatively weak, and the glass tube is relatively loose, so the internal spring needs to be replaced in time, so that the elastic force of the spring can be effectively increased. When there is mercury leakage in the gap between the switch of the mercury kettle, the screw device on the switch shaft and above should usually be tightened in time, and a certain amount of strong glue can be added around the switch shaft device. When the pressure reaches 38kpa, it needs to stay for about a minute, and if there is still mercury leakage in the gap between the switches, then it is necessary to replace the mercury bottle inside in time. When the pressure rises further to 38kPa, if the mercury column at the bottom of the sphygmomanometer device is broken, it can indicate that the internal mercury content has been obviously insufficient, and air will enter the glass tube, and effective depressurization measures need to be taken immediately, otherwise the internal mercury will emerge from the top position of the sphygmomanometer device.

Second, the sphygmomanometer cannot be inflated

1. The reason why the sphygmomanometer cannot be inflated

In the process of metrological verification, the sphygmomanometer device is easy to inflat, which is generally manifested in the fact that after the exhaust valve is closed, the pressure cannot really rise when the rubber ball is pressurized, or the pressure will rise significantly when the internal rubber ball is pinched, but the mercury column will drop immediately after letting go, which can show that the gas path of the sphygmomanometer device has leakage.

2. The sphygmomanometer cannot be inflated for fault analysis

The common strategy in the process of fault maintenance is to pressurize the sphygmomanometer device, fold and compress the rubber tube inside the mercury storage bottle to the rubber ball and the wall belt gas respectively, and then find out the specific leakage location according to the decline of mercury. The screw fastening type in the sphygmomanometer device can clearly see the rubber tube after it is opened, so after the ball is removed, the extraction type can also take out a section of rubber pipe in the exhaust valve, and then continue to rub it will see that there are cracks on both sides of the problem, and the debris in the internal cracks will be cleaned up in time, which can effectively make the air leakage fault of the sphygmomanometer device be completely solved.
Fault repair of Omron blood pressure monitor
  3. The glass tube is not vertical

1. The reason why the glass tube is not vertical

If the glass tube in the sphygmomanometer does not show a vertical phenomenon, the error of the display value verified by the sphygmomanometer device will increase significantly with the gradual increase of its internal pressure, and the positive error value of the displayed value should be 38kpa.

2. The glass tube is not vertically maintained

The glass tube inside the sphygmomanometer device can be poured forward first, and then the spring plate screw in the housing can be further loosened, and the spring plate can be gradually moved forward, and then the screw on the spring plate of the blood pressure monitor device can be tightened. And you can also tilt the glass tube to the rear, really loosen the screw on the spring plate in the sphygmomanometer, move the spring to the back, and then tighten the screw effectively, if the material of the shell is made of aluminum alloy, then the screw should not be tightened too tightly.

Fourth, the measurement results have a large error

1. The reason for the large error of the measurement results

Under normal circumstances, the staff of the measurement and verification of the sphygmomanometer device have made mistakes or unreasonable operation methods in the process of operation. The common manifestation is that the switch of the sphygmomanometer device is not closed in time after the use of the device, and after the upper cover is completely closed, there will be a part of the mercury leaking to the outside of the upper end of the sphygmomanometer glass tube. Or if the switch is turned off without a certain degree of tilting, it may eventually lead to the part of the mercury column below the zero scale line being blocked from the mercury storage bottle, and it will be retained in the value of the glass tube, and the above-mentioned mercury leakage will occur after the lid is closed. In some cases, if the lid is opened at the next operation, the remaining mercury will not be returned to the zero mark in time, and this part of the mercury will flow out of the upper part of the glass tube. Usually, the mercury storage bottle holder and the shaft of the switch and other components are more prone to mercury leakage, and when the two parts do not form a good match, mercury spraying will be triggered in the process of pressurizing the sphygmomanometer.

2. The measurement result error is large and repaired

When the above-mentioned fault occurs, usually need to replace the above-mentioned two parts in time, and in the actual working process, there are some mercury supplementation means that are relatively convenient to operate, can use the needle of the syringe to inhale the scientific, appropriate amount of mercury liquid in the spare mercury, and then open the upper cover of the sphygmomanometer device and the mercury storage bottle, insert the small hole of the tablet position at the top of the glass tube into it with a needle, and slowly inject the mercury into it, and then use the sphygmomanometer device to vibrate gently, The mercury ejected from the mercury storage bottle is fused with the newly injected mercury liquid. The final amount of mercury injected should meet the basic standard of zero error, and a smaller-diameter needle device should be used as much as possible, so that it is easier to prevent the filter from being damaged by the outside world.

More on that
The reason why the ER light is always on after the Omron PLC is powered on The reason why the ER light is always on after the Omron PLC is powered on

1. Fault phenomenon: The ERR light is always on when the PLC is powered on, and the PLC is not working. 2. Cause of failure: (1) Poor use environment leads to oxidation and corrosion of internal MPU, memory, and even substrates. (2) Due to electromagnetic interference, noise interference, static electricity, etc., the internal MPU ...

Omron PLC--CPM1A-V1 series product model Omron PLC--CPM1A-V1 series product model

1. CPM1A-10CDR-A-V1 10-point CPU unit AC100-220V, 6 points in, 4 points relay output (1A is the model code; 10 means that the total number of input and output points is 10 points, specifically 6 points of input and 4 points of output; C means CPU unit; D stands for mixed type, that is, there is a loss ...

Introduction to the functions of Omron's dual-channel intelligent fiber amplifier E3X-MZV Introduction to the functions of Omron's dual-channel intelligent fiber amplifier E3X-MZV

Dual-channel intelligent fiber amplifier E3X-MZV   Optical fiber sensors are widely used to detect the presence and absence of vibration plates, rechargeable battery pole piece winders, digital product assembly and other equipment. Omron's new fiber optic amplifier selects the functions and performance required for detection and performance.

Common problems and solutions of Omron PLC Common problems and solutions of Omron PLC

1. Omron PLC cannot be connected to the computer There are two reasons why the PLC and the computer cannot be connected: hardware and software. The hardware reasons are: the interface type does not match (such as USB, RS232, RS422, RS485, CAN, etc., both sides should be the same.) If you want to use the corresponding converter if you want to), ...

Omron MMC Health Convenience Store is once again stationed in Xi'an Omron MMC Health Convenience Store is once again stationed in Xi'an

On May 12, Omron Health Care (China) Co., Ltd. and Xi'an Yikang Pharmaceutical jointly built Xi'an Yikang Pharmaceutical Youyi West Road Store MMC Health Convenience Store (No. 105, 1st Floor, Life North District, Optical Machinery Institute, No. 121 Youyi West Road, Beilin District, Xi'an City) was officially opened and put into operation. The double ...

The correct use and precautions of Omron electronic blood pressure monitor The correct use and precautions of Omron electronic blood pressure monitor

How to use Omron blood pressure monitor 1. Rest for 10-15 minutes before using an Omron sphygmomanometer to measure blood pressure to eliminate fatigue and excitement. You must rest for 30 minutes after exercising. The indoor ambient temperature should be kept at about 20°C. Take 3-4 deep breaths before measuring your blood pressure. Lying ...

What are the selection factors of Omron relays - intermediate relays? What are the selection factors of Omron relays - intermediate relays?

1. Factors of geographical location and climate It mainly refers to factors such as altitude, ambient temperature, humidity, and electromagnetic interference. Considering the universal applicability of the control system, it must be taken into account for many years Due to the particularity of reliable operation for many years, the key parts of the device must be selected with high insulation, strong resistance...

Common troubleshooting methods for Omron relays Common troubleshooting methods for Omron relays

First, the contact point is loosened and cracked The contact point is the electrical contact part of the Omron relay to convert the load, and the contact point of some products is matched with each other by riveting pressure, and the key disadvantage is that the contact point is loose, the contact point is cracked or the error of the standard part is too large. This will jeopardize the relay ...

Omron's digital × low-carbon solutions were unveiled in the industrial automation exhibition area Omron's digital × low-carbon solutions were unveiled in the industrial automation exhibition area

The 5th China International Import Expo opened in Shanghai. Based on the ever-evolving manufacturing concept "i-Automation!" The digital and low-carbon solutions created were presented to the audience on the heavyweight platform of the Expo. At present, the curtain of global carbon neutrality has opened, and the ...

Analysis of common Omron fault codes Analysis of common Omron fault codes

1. C200H series PLC fault codes: - E0001: Undefined program error - E0002: Undefined instruction error - E0003: Missing Program Error - E0004: Data error - E0005: Pointer error - E0006: ...

Omron Switches – PCB Design Considerations for Switching Power Supplies Omron Switches – PCB Design Considerations for Switching Power Supplies

In the design of switching power supply, the physical design of the PCB board must be proper, otherwise it may cause excessive electromagnetic interference and lead to unstable power supply. 1. The design process from the schematic diagram to the PCB should be rigorous and perfect: the establishment of component parameters ...

Omron Switches – What Aspects Are Needed for Switching Power Supply Design? Omron Switches – What Aspects Are Needed for Switching Power Supply Design?

1. Wiring The design requirements of wiring are very strict in the design of switching power supplies, and they must be done well to pass the test. If the designer does not handle the wiring work in the early stage of design, there will be great safety hazards in the future electricity consumption. So here I recommend you to buy a switching electric ...

How to connect Omron plc IP address? How to connect Omron plc IP address?

1. The IP address of the computer is automatically obtained. 2. The computer is directly connected to the PLC network port. 3. Open the DirectEthernetUtility software. Software Setup Start - > All Programs - > OMRON - >Communications Middleware Utilities - > DirectE...

How to choose Omron relays? How to choose Omron relays?

1. Select the rated working current of the electromagnetic relay coil: For the DC electromagnetic relay driven by transistors or integrated circuits, the rated working current of the coil (generally 2 times the suction current) should be within the output current range of the driving circuit. 2. Select electromagnetic relay contact ...

Omron inverter maintenance failure alarm code detailed description Omron inverter maintenance failure alarm code detailed description

Omron inverter maintenance fault alarm is mainly divided into abnormal alarm and warning alarm, here we only explain the code displayed Warning display Over Torque Detection (OL2) FUNCTION code error (CE01) Registers NO. Error (CE02) Data...

The Omron CK3M/CK5M series has four major features to solve the problem of motion control The Omron CK3M/CK5M series has four major features to solve the problem of motion control

How can ultra-precision operations be realized in semiconductor and FPD manufacturing/inspection equipment, such as exposure machines, coating machines, dispensing machines, and wafer inspection equipment? How to control special mechanical mechanisms, such as those with robotic mechanisms, in the manufacture of smart devices? Machining Machine/Stamping Machine ...

Omron's digital predictive maintenance prototype enables predictive maintenance without relying on skilled workers Omron's digital predictive maintenance prototype enables predictive maintenance without relying on skilled workers

With the development of society, manufacturing sites are facing new issues, such as the lack of skilled maintenance personnel, the aging of existing equipment, and the increase in equipment accident losses due to high value-added products. To this end, Omron relies on mature automation technology to electronically transform maintenance skills ...

What are the mounting methods of Omron sensors? What are the mounting methods of Omron sensors?

The sensor can be fixed in two ways: PCB and disc mounting 1. PCB installation method In general, the primary measurement signal of the PCB-mounted sensor is relatively small. The output terminal of the sensor adopts the way of pi pin or SMD, which can be performed according to the welding process of the general device in soldering.

What are the common faults of Omron PLC? What are the common faults of Omron PLC?

1. CPU abnormality: When the CPU abnormality alarms, check the devices connected to the internal bus of the CPU unit. The specific method is to replace the units that may cause failures in turn, find out the faulty units, and deal with them accordingly. 2. Memory abnormality: When the memory is abnormal alarm, if it is a program ...

【Omron industry case】Predictive maintenance of smoke machine in automobile welding workshop 【Omron industry case】Predictive maintenance of smoke machine in automobile welding workshop

Process introduction Automobile welding is the second of the four major processes in vehicle production, which mainly assembles stamped body components into a complete body. In the welding process, spot welding and arc welding of robots are the mainstay, and there will be a station for visual inspection of the welding status in the middle.   Make...

CATEGORIES BYPASS
Customer Service Center

Online Consultation:QQ


ContactContact

Contact: Manager Huang

Contact QQ: 3271883383

Contact number: 13522565663


Scan the code to add WeChat (please save the picture first on the mobile phone)

working hoursworking hours

Weekdays: 9:00-17:00

Holidays: Only emergencies are handled

Contact us

Contact us

Contact number QQ consultation
QQ consultation

3271883383

Company address
Back to top