Introduction: E+H/Endershaus Modular RTD Thermometer TR25, the thermometer is mainly used in the chemical industry, but also for other industrial applications. Head-mounted transmitter-mounted devices are ready-to-use, enhancing measurement accuracy and reliability compared to directly connected sensors. A wide range of process connections, sizes, and materials provide application flexibility.
Features:
Modular design with standard junction boxes in accordance with DIN EN 50446 and customer-specified immersion lengths for maximum flexibility
The cores are highly compatible and designed in accordance with DIN 43772
Short tapered ends for faster reaction times
The head transmitter is easy to select: analogue outputs from 4 to 20 mA, HART®, PROFIBUS® PA, or FF fieldbus
Product parameters
Measurement accuracy
Cl. A, according to IEC 60751 standard
Cl. AA, according to IEC 60751
Response time
Depends on the settings
t50 = 2 s
t90 = 5 s
Maximum process pressure (static pressure)
At 20 °C: 40 bar (580 psi)
Operating temperature range
PT 100:
-200...600 °C
(-328...1.112 °F)
The maximum insertion depth required
up to 10,000.00 mm (393.70'')
Our customer-centric service concept: the pursuit of a full range of services is a high-quality, high-standard service, the pursuit of every detail of the service process. It is impossible for interests to blossom into brilliant flowers from lies. Only when every process, every action and every detail of the service is professional and careful, can we meet or even exceed the expectations of customers, so that customers can be satisfied and recognize us. Honest, professional, affordable, and fast, so you can rest assured. "Honest and professional service, affordable price, continuous improvement" is our service philosophy.